Omega 24 User Manual

Browse online or download User Manual for Thermometers Omega 24. Omega 24 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - CL23A, 24, 25, 26, 27

hhtp://www.omega.com e-mail:[email protected] CL23A, 24, 25, 26, 27Digital Calibrator/Thermometer

Page 2

WARNING Do not touch a temperature probe sheath when measuring excessively high or low temperatures, or toxic substances. WARNING Do not atte

Page 3 - TABLE OF CONTENTS

CAUTION Keep inputs electrically isolated from outputs. Do not con- nect an external voltage source between the two connectors, and a

Page 4

2. Remove the old battery. 3. Place the new battery in the battery compartment. Be sure to observe proper polarity. 4. Re-install the battery co

Page 5

OPERATING INSTRUCTIONS 1. DIAGNOSTICS AND ERROR MESSAGES Unit self-diagnostics provide fault condition readouts which are described below: PROBLE

Page 6

Figure 2. Keypad Figure 3. Display Test If the unit is turning on for the first time after a new battery is installed

Page 7 - UNIQUE SPECIFICATIONS

METER: (unit functioning as a thermometer) K: (K-type thermocouple) °F: (Fahrenheit readings) OPR: (Instrument is o

Page 8 - MANUAL ADDENDA

6. Select display resolution with 0.0.1 key. NOTE: Set-up is retained during power-off. Hold: Meter readings can be put on hold at any time by d

Page 9 - SAFETY PRECAUTIONS

Figure 6.4-Wire Resistance Measurement Lead resistance errors are compensated for in 3-wire configurations. However full compensation

Page 10

To Save Readings: 1. While in the METER OPR mode, depress the STORE key. Note that the OPR annunciator is replaced by STO. 2. Next depress one

Page 11

eg.RECALL 3 RECALL 5 RECALL 2 etc. 4. To return the meter to the operate mode, depress OPR after recalling the last reading. A

Page 12 - BATTERY

Servicing North America USA: One Omega Drive, Box 4047 ISO 9001 Certified Stamford, CT 06907-0047 Tel: (203) 359-1660 FAX: (203

Page 13 - OPERATING INSTRUCTIONS

6. Select the resolution desired (0.0.1 key). NOTE: With 1° resolution, one less key stroke will be required to enter calibrator ou

Page 15

The same performance tradeoffs apply to 2, 3 and 4-wire simulations as to 2, 3, and 4-wire measurements. Refer to RTD, Thermistor and Ohms

Page 16 - (RTD, Thermistor, etc.)

To Recall Outputs: 1. While in the CALIB OPR mode, set-ups saved earlier in memory can be recalled to program the calibrator output. First depres

Page 17

9. STEP FUNCTION When operating in the CALIBRATOR mode, the STEP key will sequentially recall data stored in memory (up to 10 steps). At each ste

Page 18

SERVICE INFORMATION WARNING All service information is intended for qualified electronic mainte- nance personnel only. 1. CALIBRATION PROCEDU

Page 19

NOTE: For calibrators with the older, single-jumper configuration, refer to calibration procedure beginning on page 39. 4. Install calibr

Page 20

Calibration Adjustments: NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure

Page 21

*exclusive of noise. **on EC models, G replaced by L.Calibrate L per J. 26

Page 23

TABLE OF CONTENTS General Information ... 2 Common Specifications.

Page 24

B. RESISTANCE CALIBRATION NOTE: Thermocouple and resistance calibrations are independent. However, potentiometer P1 is used in both the

Page 25 - WARNING

Figure 14. Resistance Calibration Setup 29

Page 26

Calibration Adjustments: NOTE: Do not deviate from the calibration adjustment sequence that follows. This will ensure that adjust

Page 27

* exclusive of noise 31

Page 28

Table 6: Calibration of Calibrator-Mode Resistance Functions (RTD-Thermistor) *exclusive of noise 2. CALIBRATION VERIFICATION Calibration i

Page 29

Model Software Revision CL23A 6.7 CL24 7.7

Page 30

4. Select type K thermocouple ("K" on display) by pressing the SENSOR SELECT function button until type K is reached. 5. Select the te

Page 31

19. Press the SENSOR SELECT function button. The sensor type will change to type T. The "J" will disappear from the display and a &qu

Page 32

32. Press the SENSOR SELECT function button. The sensor type will change to type C. A "C" will appear momentarily in the center of

Page 33

3. Select the calibrator mode ("CALIB" on display) by pressing the CALIB/METER function button. 4. Select type K thermocouple ("K&q

Page 34

GENERAL INFORMATION This manual provides operating instructions and maintenance information for five temperature instruments. These instruments a

Page 36

CALIBRATION PROCEDURES Test Equipment Required: 1. Thermocouple simulator (OMEGA CL25, or equivalent). 2. Microvolt null-meter (Keithley 155,

Page 37

• With J1 removed, the digital-filters in the U.U.T. are dis- abled. This is to speed up U.U.T. response to calibra- tion adjus

Page 38

23. Select Type-E thermocouple. 24. Turn off U.U.T. 25. Replace T calibration-cable with E calibration-cable. 26. Turn on U.U.T. Allow 2-3 minu

Page 39

45. Replace J calibration-cable with T calibration-cable. 46. Turn on U.U.T. Allow 2-3 minutes for thermal stabilization. 47. Set both the U.U

Page 40

NOTES 43

Page 43

RTD; ±0.2°F (rdg> 50°F) ±0.5°F (rdg < - 50°F) ±0.2°F ±0.04% rdg (meter mode) THERMISTOR;

Page 44

KEYPAD: 12 momentary switches with tactile feedback select; • On/Off • Change/Enter • Clear/-0 • Sensor Select/1 • °F/°C/2

Page 45

BATTERY LIFE, CONTINUOUS: 10 hrs. typical, alkaline; 3 hrs typical, Ni-Cd (rechargeable). BATTERY INDICATOR: Display indicates BAT when less than 10

Page 46

ACCURACY: (18°C to 28°C ambient, 2 years): ±0.5 °F (rdg > -50°F). ±0.04% rdg (rdg > 1250°F)

Page 47

UNPACKING AND INSPECTION Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, pl

Related models: 25 | 27 | 26 |

Comments to this Manuals

No comments